Relative Analgesia Machine Features:
Oxygen Fail-Safe: if the oxygen supply is interrupted or inadvertently cut off, the RAŽ will immediately discontinue the flow of nitrous oxide.
Minimum Flow of 3 LPM Oxygen: the RA will not provide oxygen flow below 3 LPM.
Automatic Air Intake Valve: if the contents
of the breathing bag are depleted, the air intake valve automatically opens
providing room air to the patient.
Non-Rebreathing Circuit:
a check valve in the mixed gas outlet protects the patient from rebreathing
exhaled gases.
20 LPM Oxygen Flush:
an integrated oxygen flush button conveniently located on the front of the
RA provides an immediate 20 LPM oxygen flow to the patient.
Auxiliary Oxygen Outlet: an outlet is provided for the attachment of the optional "3-in-1 " Demand Valve Resuscitator.
D.I.S.S. Gas Connections: gas connections on the RA conform to the industry standard diameter indexed safety system (D.I.S.S.)
Finger-Guard On/Off Switch: an anodized aluminum finger guard prevents accidental activation/deactivation of the single on/off switch.
Positive Stop Needle Valves: sensitive, easily adjustable valves control the flow rate of each gas separately.